Summary
Type: SPE
Application #: 300530
Catalog#: 730073 - SPE Cartridges, Chromabond, Silica, 3ml, PP, 500mg, 50/pk
Detection:
Conditions: Column conditioning:
2 ml n-hexane / acetone (8 : 2, v/v)
Sample application:
Slowly force or aspirate 1 ml sample through the column
Column washing:
2 ml n-hexane / acetone (8 : 2, v/v); dry 3 min under vacuum
Elution:
2 x 500 μl methanol
Authors: Macherey-Nagel
Source: LC department